AMETEK 933 — Alarm Codes (Quick Actions)

Common AMETEK 933 alarm conditions with recommended first checks. Use decision trees for step-by-step isolation.

Field tip: For oxygen analyzers, solve flow/leaks/moisture before touching calibration or sensor replacement.

Flow / Sample System

  • Low / No Flow — confirm sample supply, filter blockage, regulator setpoint, pump operation
  • Flow unstable — regulator hunting, leaks, pump weak, condensation
  • High DP / Filter blocked — replace/clean filter, check sample particulate

Reading Quality

  • Unstable reading — moisture, leaks, unstable flow, loose wiring, grounding
  • Slow response — long lines, restrictions, weak pump, moisture separation issue

High O₂ Bias

  • O₂ reading too high — air ingress/leaks, cracked tubing, loose fittings, sample dilution
  • Span seems high but process stable — verify calibration gas and sample integrity

Calibration / Sensor Health

  • Calibration failure — wrong gas, leaking cal manifold, insufficient warm-up, sensor aging
  • Zero/Span won’t stabilize — flow/leaks/moisture first, then sensor