Decision Tree: High O₂ Bias (Air Ingress / Dilution)

Symptom: O₂ reads higher than expected; process stable but analyzer high.

Key idea: Most high bias is from air leaks into a negative-pressure or low-pressure sample system.

Isolation steps

1) Compare reading on calibration span gas. If span is correct, the sensor is likely OK → suspect sample leak/dilution.
2) Leak check sample system:
  • Check fittings from probe/sample point to analyzer
  • Check filter bowls, drain points, unions
  • Check any flexible tubing or cracked lines
3) Check sample pressure regime:
  • If sample line is under suction (pump pulling), small leaks pull air in → high O₂.
  • Stabilize with correct regulator/flow and minimize suction leakage points.
4) Confirm sample integrity:
  • Ensure sample point not drawing ambient air (bad probe sealing)
  • Confirm purge/vent lines not open to atmosphere incorrectly
5) If no leak found:
  • Verify calibration gas values and regulator accuracy
  • Check sensor drift/aging if both process and cal are high
Field tip: If the analyzer is high only when sample flow is low, suspect a leak at a fitting or drain point. Fix flow first and re-check bias.