Decision Tree: Poor Repeatability

Symptom: repeat injections vary; unstable component readings or peak areas.

Field tip: If peak area varies → injection/sample issue. If peak timing varies → flow/temperature issue.

Step-by-step isolation

1) Check sample pressure stability and fast loop flow. Unstable sample pressure causes volume variation.
2) Inspect sample conditioning: filters, regulators, condensation, liquid carryover, heat tracing.
3) Valve/injection loop check:
  • Listen/confirm valve switching sequence is consistent.
  • Inspect valve wear, sticking, solenoid response.
  • Check for contamination in sample loop.
4) Carrier flow stability: verify pressure/flow does not fluctuate (no regulator hunting).
5) Detector stability: if baseline is noisy, fix baseline first (see baseline tree).
6) Validation:
  • Run a calibration standard 3 times. If standard repeats OK but process does not → process/sample variability.
  • If standard also varies → instrument/sample system problem.
7) Corrective actions:
  • Stabilize sample regulators and repair conditioning system.
  • Service/replace injection valve components if sticking or worn.
  • Replace contaminated filters and verify with repeats.