Yokogawa – Alarm Decision Trees
Step-by-step alarm troubleshooting flows written for analyzer technicians. Follow in sequence — fix sampling & utilities before electronics.
Golden rule: If readings are unstable, drifting, or wrong —
check sample condition, pressure/flow, purge/reference air before calibration or parts replacement.
Select Analyzer
ZR22G – Zirconia O₂ Analyzer
Heater issues, drift, calibration failures, no response.
OX400 – Sampled-Gas O₂ Analyzer
Model-specific documentation notes for unstable readings, calibration and outputs.
GC8000 – Process Gas Chromatograph
Carrier pressure, oven readiness, valve switching, detector faults.
Technical references
Updated against manufacturer documentation. Match the document to the installed model and options.
- Yokogawa: OX400 Low Concentration Zirconia Oxygen Analyzer — GS 11M10B01-01E, 3rd edition, March 2011; amended pages June 2017.