Field rule: If temperature/flow is not correct, readings are not valid.
Fix sample integrity before suspecting analyzer electronics.
Quick Checks (Do these first)
- Safety: confirm SRU area safety, permits, and isolation rules.
- Sample temperature: verify heater setpoint vs actual; check for cold spots.
- Sample flow: confirm flow indication; check restrictions.
- Leaks: check sample system for air ingress (especially at fittings/manifolds).
- Filters: check filter ΔP / condition; replace if fouled.
- Calibration gases: verify correct cylinder and regulator stability.
Symptom: Low / Zero Reading
- Check: sample flow low? filter clogged? probe plugged?
- Check: sample temperature low? condensation/sulfur dropout likely.
- Check: cal valve stuck in zero? manifold leakage?
- Likely cause: probe/line plugging, sulfur deposition, cold spot, valve fault.
- Fix: restore heat, clean/replace filters, verify blowback (if available), verify valves return to process.
Symptom: High Sulfur Reading / Sudden Step Change
- Check: validate with span gas (does it track expected value?).
- Check: process upset? compare SRU parameters and any redundant analyzer.
- Check: air leakage can bias reading depending on measurement method/config.
- Likely cause: real SRU upset, calibration drift, sample contamination.
- Fix: confirm calibration, inspect sample system, correlate with DCS trends.
Symptom: Unstable / Noisy Reading
- Check: sample pressure fluctuation (regulator hunting, restriction).
- Check: water/moisture ingress; confirm heater temps and drains.
- Check: grounding/shielding; inspect cable routing near VFD/motors.
- Likely cause: sample instability, moisture, electrical noise, detector contamination.
- Fix: stabilize pressure, correct heat, improve grounding/shielding, clean sample path.
Symptom: Slow Response / Sluggish Trend
- Check: long sample line, dead volume, partially plugged probe/filter.
- Check: sulfur deposition (common SRU) causing adsorption and delay.
- Likely cause: restrictions, cold spots, internal fouling.
- Fix: clean/replace probe and filters, eliminate cold spots, verify heater performance.
Symptom: Calibration Failure
- Check: correct gas cylinder + concentration? regulator stable?
- Check: cal line leaks? cal valve passing or stuck?
- Check: sample system still contaminated/sulfur dropout?
- Likely cause: wrong gas, leaking manifold, insufficient warm-up, sample contamination.
- Fix: verify cylinder, leak test manifold, restore proper temperature/flow, repeat calibration.
Symptom: Analyzer Offline / No Output
- Check: power supply, fuses, UPS, internal breaker.
- Check: I/O wiring to DCS (4–20 mA loop power, AO range, scaling).
- Check: network/Modbus settings if digital communication used.
- Likely cause: power fault, controller fault, loop wiring issue.
- Fix: restore power, verify loop, check configuration, reboot only after confirming stable utilities.
SRU Field Best Practices
Tail gas analyzers fail most often because of sample system design/maintenance. Use this guide to prevent recurring issues.