Servomex 4900 — Multigas Process Analyzer
Continuous multigas analyzer using paramagnetic oxygen and NDIR infrared measurement technologies. Widely used in hydrogen plants, refineries, SMRs, and air separation units.
Technician rule:
If readings are unstable, always check
sample conditioning, pressure stability, and moisture
before calibration or sensor replacement.
Working Principle
The Servomex 4900 uses two independent physical principles for continuous gas measurement.
- Paramagnetic O₂: Oxygen molecules are attracted into a magnetic field, generating a signal proportional to concentration.
- NDIR (CO / CO₂): Infrared absorption at specific wavelengths determines gas concentration.
Main Components
- Sample conditioning: filters, regulators, drains.
- Flow control: capillaries / flow restrictors.
- Measurement modules: paramagnetic O₂ sensor, IR cell.
- Electronics: signal conditioning & processing.
- Controller & comms: display, I/O, Ethernet.
Calibration Procedure
1) Confirm analyzer is stable (no flow alarms).
2) Verify sample & calibration gas pressure.
3) Apply zero gas and allow stabilization.
4) Apply span gas and verify response.
5) Return to process and confirm DCS values.
Major Troubleshooting
- Unstable readings: pressure hunting, wet sample, leaks.
- Slow response: blocked filters or restricted flow.
- High drift: sensor ageing or contamination.
- No DCS update: comms, scaling, or quality-bit issues.
Maintenance (Field Practice)
- Daily: Check sample flow, pressure, and alarms.
- Weekly: Inspect filters, drains, and fittings.
- Monthly: Leak test sample system, verify calibration.
- Annually: Sensor health check and validation.
Poor maintenance causes more analyzer faults than electronics failure.
Discussion
Share Servomex 4900 field experience, calibration issues, sensor failures, and plant-specific fixes. (Requires GitHub login.)