Decision Tree: Communication / DCS

← Back to Alarms Hub Alarm Codes → Back to MAXUM II
Goal: restore stable updates to DCS/PLC/Historians (values + quality + update timing).
1) Confirm analyzer is running analysis normally (latest run time updating).
2) Check physical/network basics: link lights, correct IP/subnet/gateway, ping test.
3) Verify protocol settings: Modbus TCP / OPC / serial gateway match DCS config.
4) Verify mapping/scaling: register list, data types, scaling, and quality/status mapping.
5) Align DCS polling interval with GC cycle time (avoid timeouts).

Quick Symptoms → Likely Cause

All values stale: IP/network, service down, wrong port.

Values update but wrong: mapping/register offset, scaling, wrong data type.

Bad quality only: quality/status mapping, handshake, polling mismatch.