Calibration Best Practices (Field Proven)
Calibration does not fix unstable analyzers. This guide explains when to calibrate, when NOT to calibrate, and how to avoid false drift.
Golden rule: Never calibrate during alarms, instability,
or abnormal process conditions.
Before Calibration (Mandatory)
- No active flow, pressure, or system alarms
- Stable readings for at least 30–60 minutes
- Correct calibration gas connected
- Sample system verified dry and leak-free
Calibration Gas Handling
- Use certified gas with valid expiry date
- Confirm gas concentration matches analyzer range
- Do not reuse partially empty cylinders
- Use proper regulators (no needle valves)
Zero & Span Application
- Allow full stabilization time (never rush)
- Confirm zero returns correctly after span
- Do not “chase” readings with repeated adjustments
Verification (Most Skipped Step)
- Apply verification gas after calibration
- Confirm result within acceptance limits
- Document results before returning to process
Common Calibration Mistakes
- Calibrating during pressure hunting
- Ignoring moisture in sample lines
- Using wrong gas or wrong concentration
- Blaming sensor instead of sampling
Field truth: Most “drift” complaints are caused by
unstable sampling — not sensor ageing.