Sample path (think in blocks)

Stack/Process → Probe → Heated Line → Filter(s) → Pump/Flow Control → Conditioning (chiller/dryer) → Analyzer → Vent/Return

Most common field failures

Probe Plugging

Low flow, slow response, unstable readings. Usually dust/ash/salts buildup.

Cold Spots / Heater Failure

Condensation forms, causing drift, noise, corrosion, and repeated calibration failure.

Filter Restriction

Pressure drop increases, pump works harder, response time becomes slow.

Leaks (Air Ingress)

False O₂, dilution of gas analyzers, unstable baseline. Leak test is mandatory.

Technician troubleshooting method (sample system)

  1. Confirm flow at analyzer inlet (not just pump running)
  2. Check filters and DP (if available)
  3. Inspect chiller temperature and drain
  4. Check heated line temperature and continuity
  5. Perform leak test (calibration manifold + sample path)
  6. Verify stable readings using calibration gas
Rule: If response is slow, fix flow restriction first. If readings drift, suspect condensation/leaks before electronics.